Intraoperative Magnetic Resonance Imaging (iMRI) mishaps – Troubleshooting an unsafe object attached to the scanner


Department of Neuroanaesthesia, All India Institute of Medical Sciences, New Delhi, India

Date of Submission01-Jul-2020
Date of Decision15-Jul-2020
Date of Acceptance28-Aug-2020
Date of Web Publication20-Mar-2021

Correspondence Address:
Ashish Bindra
Room No 710, Neuroanaesthesia and Critical Care, Neurosciences Centre, AIIMS, New Delhi
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ija.IJA_870_20

Rights and Permissions

How to cite this article:
Sameera V, Bindra A, Rath GP. Intraoperative Magnetic Resonance Imaging (iMRI) mishaps – Troubleshooting an unsafe object attached to the scanner. Indian J Anaesth 2021;65:50-1

How to cite this URL:
Sameera V, Bindra A, Rath GP. Intraoperative Magnetic Resonance Imaging (iMRI) mishaps – Troubleshooting an unsafe object attached to the scanner. Indian J Anaesth [serial online] 2021 [cited 2021 Apr 10];65:50-1. Available from: https://www.ijaweb.org/text.asp?2021/65/13/50/311605



Sir,

The unfortunate incident of 2001, where an oxygen cylinder accidentally brought into the vicinity of a magnetic resonance (MR) scanner resulted in a fatal injury to a 6-year-old boy, has impressed upon the world the grave consequences of projectile injuries.[1] Despite strict regulations and multiple guidelines near-misses continue to be reported. Incorporation of magnetic resonance imaging (MRI) into the intraoperative environment has added to neurosurgical armamentarium but has also potentially increased the scope for errors. We report an unsafe event in the intraoperative MRI (iMRI) operation room (OR). Our OR support worker was replacing fresh linen in the iMRI OR while preparing it for the second case. Due to the force of habit as in other ORs, on that fateful day, he entered the OR in a hurry with a monitor still in his hand. It led to the pulling of the transport monitor to the magnetic console [Figure 1]. This is how, despite many safety precautions and protocols in place, an MR unsafe transport monitor found its way into BrainSuite© iMRI. (Brainlab AG, Germany). BrainSuite iMRI holds both scanner and MR compatible table in the same room. Surgery is conducted outside the 5G line and the table is rotated 180 degrees to access gantry for a scan. Presence of theatre support workers doing duty in both iMRI as well as a normal OR resulted in an error of judgment. Fortunately, there were no physical injuries but the consequences of the event were, nevertheless, severe.
Figure 1: MR unsafe monitor attached to the bore of MRI scanner

Click here to view


iMRI poses unique challenges which are unlike the diagnostic MRI suites. iMRI lasts for longer and complex therapeutic procedures, requiring more manpower and equipment (the navigation system, microscopes, endoscopes, drills, surgical instruments, neuromonitoring, and anaesthetic equipment) with limited options to handpick MR safe/conditional devices. These facilities are manned by a team of nurses, technologists, and cleaning staff who work in both iMRI ORs as well as normal ORs. iMRI setups are largely proclaimed safe. Tan et al.[2] documented 43 MR undesirable incidents in 19 months of iMRI usage, none of which resulted in injury to the patient. However, 32.6% of these incidents were attributed to human errors and were preventable.

Troubleshooting of an undesired object attached to a magnetic bore is worth discussing [Figure 2]. Technical support is not readily available in iMRI, hence steps of troubleshooting, as well as contact details of engineers, should be well displayed. The pulling of objects stuck to the MRI is ill-advised. The magnetic force of a 3 Tesla magnet bore is 60000 times of the earth's gravitational field.[3] Hence, it is near impossible to pull an object weighing a few kilograms (in our case, the monitor) against such force. Pulling might cause objects to fall back onto the scanner with force and result in damage both to the scanner and the attempting personnel (if trapped between the scanner and the object to be pulled). Objects may shatter during the pulling attempt; internal ferromagnetic components might disperse and result in flying objects causing injury. Pulling should only be reserved for very small objects. The Quench button should be used only in an absolute (life/limb-threatening) emergency. Quench quickly shuts off the magnetic pull, but the economic burden of the process is huge due to the non-recovery of helium. The alternative is to slowly ramp down the magnetic field strength to the level sufficient to safely retrieve the object. Ramping down is associated with less helium loss and expenditure compared to a quench. However, it is a slow process (2-3 days), and ramping up takes time potentially rendering the MRI scanner unusable for a few days.
Figure 2: Steps of troubleshooting when object gets pulled into the scanner

Click here to view


Strict enforcement of checklists and protocols is the key to avoid mishaps.[4],[5] A dedicated person with the sole responsibility to oversee screening and restrict entry, mandatory tutorials at regular intervals (to refresh knowledge and reiterate safe practices) and pocketless scrubs for personnel entering MR suite are some of the suggestions that could be implemented in the Indian scenario.[6]

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Gilk T, Hsdq MA, Latino RJ. MRI Safety 10 Years Later: what can we learn from the accident that killed Michael Colombini? Patient Safety & Quality Healthcare 2011;8:22-3. Available from: https://www.psqh.com/analysis/mri-safety-10-years-later/. [Last cited on 2020 May 15].  Back to cited text no. 1
    
2.
Tan JKT, Tan TK, Goh JPS, Ghadiali NF. Prospective review of safety incidents reported in the iMRI OT (Intraoperative Magnetic Resonance Imaging Operating Theatre). Proc Singap Healthc 2014;23:273-81.  Back to cited text no. 2
    
3.
Panych LP, Madore B. The physics of MRI safety. J Magn Reson Imaging 2018;47:28-43.  Back to cited text no. 3
    
4.
Manohar N, Mohapatra D, Balasubramaniam A, Rao K, Srinivas D, Chakrabarti D. Setting up workflow of an intraoperative MRI unit: A single-centre experience of first 53 cases. J Neuroanaesth Crit Care 2018;05:177-83.  Back to cited text no. 4
    
5.
Gandhe RU, Bhave CP. Intraoperative magnetic resonance imaging for neurosurgery – An anaesthesiologist's challenge. Indian J Anaesth 2018;62:411-7.  Back to cited text no. 5
[PUBMED]  [Full text]  
6.
Wilson SR, Shinde S, Appleby I, Boscoe M, Conway D, Dryden C, et al. Guidelines for the safe provision of anaesthesia in magnetic resonance units 2019: Guidelines from the Association of Anaesthetists and the Neuro Anaesthesia and Critical Care Society of Great Britain and Ireland. Anaesthesia 2019;74:638-50.  Back to cited text no. 6
    


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed194    
    Printed0    
    Emailed0    
    PDF Downloaded58    
    Comments [Add]    

Recommend this journal