|
|
ORIGINAL ARTICLE |
|
Year : 2018 | Volume
: 62
| Issue : 2 | Page : 103-108 |
|
|
Comparison of the ProSeal laryngeal mask airway with the I-Gel™ in the different head-and-neck positions in anaesthetised paralysed children: A randomised controlled trial
Gargi Banerjee1, Divya Jain1, Indu Bala1, Komal Gandhi1, Ram Samujh2
1 Department of Anaesthesia and Intensive Care, Post Graduate Institute of Medical Education and Research, Chandigarh, India 2 Department of Paediatric Surgery, Post Graduate Institute of Medical Education and Research, Chandigarh, India
Date of Web Publication | 12-Feb-2018 |
Correspondence Address: Dr. Divya Jain Department of Anaesthesia and Intensive Care, Post Graduate Institute of Medical Education and Research, H. No-3034, Sector 21-D, Chandigarh - 160 022 India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/ija.IJA_594_17
Background and Aims: Head and neck movements alter the shape of the pharynx, resulting in changes in the oropharyngeal leaking pressures and ventilation with supragottic airway devices. We compared the effect of the different head-and-neck positions on the oropharyngeal leak pressures and ventilation with the I-Gel™ and ProSeal™ laryngeal mask airway (PLMA) in anaesthetised paralysed children. Methods: A total of 70 children were randomly assigned to receive PLMA (n = 35) or I-Gel™ (n = 35) for airway management. Oropharyngeal leak pressure in maximum flexion, maximum extension and the neutral position was taken as the primary outcome. Peak inspiratory pressures (PIPs), expired tidal volume, ventilation score and fibreoptic grading were also assessed. Results: No significant difference was noted in oropharyngeal leak pressures of PLMA and I-Gel™ during neutral (P = 0.34), flexion (P = 0.46) or extension (P = 0.18). PIPs mean (standard deviation [SD]) were significantly higher (17.7 [4.03] vs. 14.6 [2.4] cm H2O, P = 0.002) and expired tidal volume mean [SD] was significantly lower (5.5 [1.6] vs. 6.9 [2] ml/kg, P = 0.0017) with I-Gel™ compared to PLMA. Fibreoptic grading and ventilation score were comparable in both the groups in all the three head-and-neck positions. Conclusion: PLMA and I-Gel™, both recorded similar oropharyngeal leaking pressures in all the three head-and-neck positions. However, higher peak pressures and lower expired tidal volume in maximum flexion of the neck while ventilating with I-Gel may warrant caution and future evaluation.
Keywords: Airway-laryngeal mask airway, anaesthesia-paediatrics, position-head-and-neck
How to cite this article: Banerjee G, Jain D, Bala I, Gandhi K, Samujh R. Comparison of the ProSeal laryngeal mask airway with the I-Gel™ in the different head-and-neck positions in anaesthetised paralysed children: A randomised controlled trial. Indian J Anaesth 2018;62:103-8 |
How to cite this URL: Banerjee G, Jain D, Bala I, Gandhi K, Samujh R. Comparison of the ProSeal laryngeal mask airway with the I-Gel™ in the different head-and-neck positions in anaesthetised paralysed children: A randomised controlled trial. Indian J Anaesth [serial online] 2018 [cited 2021 Jan 21];62:103-8. Available from: https://www.ijaweb.org/text.asp?2018/62/2/103/225221 |
Introduction | |  |
Nowadays, supraglottic airway devices (SADs) are being used in various surgeries such as palatoplasty or burn contracture release requiring different head-and-neck positions in children.[1],[2] These movements alter the shape of the pharynx, resulting in changes in the oropharyngeal leaking pressures (OPLP) and ventilation.[3]
To date, the ProSeal™ laryngeal mask airway (PLMA) is considered the benchmark among second-generation SADs [4] The superiority of PLMA over other SADs in adults is attributed to the presence of the posterior cuff and increased the bulk of the PLMA mask resulting in increased oesophageal and pharyngeal seal. I-Gel™ (Intersurgical Ltd., Wokingham, Berkshire, UK) is a cuffless, single-use second-generation SAD made of a soft gel-like thermoplastic elastomer (styrene butadiene ethylene-styrene) which provides an excellent perilaryngeal seal.[5] Due to its potential advantages, I-Gel™ has gained widespread popularity among paediatric anaesthetists.
We were interested to know whether PLMA would be able to provide better OPLP and ventilation compared to I-Gel™ in the different head-and-neck positions in children.
Therefore, we designed this prospective randomised trial to evaluate and compare the performance of PLMA and I-Gel™ in the different head-and-neck positions in anaesthetised paralysed children. The primary outcome was the OPLP of the two study devices in maximum flexion, extension and neutral position. Secondary outcomes included the peak inspiratory pressure (PIP), expired tidal volume, ventilation scoring and grading of fibreoptic view.
Methods | |  |
The study was conducted from June 2014 to December 2015 in the advanced paediatric centre of a tertiary care hospital after obtaining approval from the Institutional Ethics Committee (NK/1606/MD/10039-40). Written informed consent was obtained from the parents/guardians of 70 American Society of Anaesthesiologists Grade I or II children of either gender weighing between 10 and 30 kg enrolled in the study. Children having upper respiratory tract infection on the day of surgery, anticipated difficult airway or children at increased risk of aspiration such as gastro-oesophageal reflux disease, non-fasting status, hiatus hernia, lung diseases and limited head-neck movements were excluded from the study.
Children were fasted 6 h for solid food, 4 h for milk and 2 h for clear water. Monitoring included pulse oximetry, non-invasive arterial blood pressure, electrocardiography and capnography (Aestiva 5™ 7900, Datex Ohme2 orda, Madison, USA) was done. General anaesthesia was induced with 100% oxygen with sevoflurane (6%–8%). Injection fentanyl 2 μg/kg intravenously (IV) was given an analgesic and injection atracurium 0.5 mg/kg IV was administered for muscle relaxation. Children were randomised to have inserted either I-Gel™ size 2 or 2.5 or PLMA size 2 or 2.5 using computer-generated randomisation chart (http://www.randomization.com). The allocation was concealed in opaque envelopes that were opened only before the start of anaesthesia. The appropriate sized allocated device was inserted following manufacturer's recommendations. An introducer was used to insert appropriate sized PLMA.
Effective ventilation of the device was defined as bilateral chest movements on gentle manual ventilation and square shaped capnograph trace. A maximum of two insertion attempts were allowed before considering it as failure to insert the device. Endotracheal intubation was performed in case failure to insert the device satisfactorily or if there was displacement of the device resulting in an inability to ventilate.
After successful insertion of the study device, the patient's lungs were ventilated for 3 min with a tidal volume of 10 ml/kg and respiratory rate of 12–20/min according to patient's age in a neutral position. The neutral position was made by aligning the external ear canal and the superior orbital margin of the eye in a vertical plane to the horizontal operating table. The OPLP, PIP, expired tidal volume, ventilation scoring and fibreoptic grading were recorded in the neutral position. Thereafter, the position of the head-and-neck was changed to maximum extension (45° from neutral) or maximum flexion (45° from neutral) in random order. The degree angle at each position was measured using a goniometer. After 30–60 s of each position change the OPLP, PIP, expired tidal volume, ventilation scoring and fibreoptic grading were recorded.
OPLP was measured using closing the expiratory valve of the circle system at a fixed gas flow of 3 L/min and recording the airway pressure at which the dial on a calibrated aneroid manometer reached equilibrium. The maximum airway pressures were not allowed to exceed 30 cm H2O. The oropharyngeal leak was determined either at the mouth (audible), the stomach (epigastric auscultation) or at the drainage tube (bubbling of lubricant placed on the proximal end of the drainage tube).[6]
Adequacy of ventilation was assessed based on three criteria: (1) No leakage with an airway pressure of 15 cm H2O, (2) Bilateral chest excursion with a PIP of 20 cm H2O and (3) Square wave capnograph. Each point was given a score of either 0 or 1 point. Thus, if all three criteria were satisfied, the ventilation score was 3.[7]
Fibreoptic grading was determined by passing a 3.5 mm fibreoptic scope (I-view™ scope, VBM India Co., New Delhi, India) through the airway tube to a position 1 cm proximal to the end of the tube. Fibreoptic scoring was done based on scores used by Okuda et al. (4 - <1/3 view covered with epiglottis, 3-1/3–2/3 view covered with epiglottis, 2 - >2/3 view covered with epiglottis, 1-completely covered with epiglottis but having an adequate function).[8]
At the end of the surgery, the device was removed after complete reversal of neuromuscular blockade. Complications with device insertion, maintenance or removal such as airway reflex activation (coughing, laryngospasm or bronchospasm), desaturation (SpO2 <90%) or blood staining on the device after removal were noted.
Sample size calculation was determined from a pilot study on ten patients. Assuming the largest difference in mean OPLP (standard deviation [SD]) cm H2O between the different positions to be 4 (4.2) cm H2O within the two study devices, namely PLMA and I-Gel™, we calculated the sample size of 31 with an alpha error of 0.05 and power of 0.8. To account for dropouts or study failures, 35 children were enrolled in each group.
Data were presented as the difference of mean (SD), median (range) or an absolute number of patients. Intragroup comparisons of OPLP, PIP and expired tidal volume were done using repeated measures analysis of variance. Kruskal–Wallis test was applied for comparing fibreoptic scoring, and Friedman's (non-parametric) test was used to compare ventilation score. Data between the two groups were analysed using the unpaired t-test and Mann–Whitney test. For the analysis of nominal data, we used Chi-square analysis or Fisher's exact test. SPSS software Version 22 (IBM Inc., NY, USA) was used for statistical analysis. P < 0.05 was considered statistically significant.
Results | |  |
Out of 100 children assessed for eligibility, 70 children were randomised according to the study protocol. [Figure 1] shows the consort diagram of the enrolled patients. All the children who were randomised completed the trial and data obtained from them were analysed.
The patient characteristics are presented in [Table 1]. PLMA and I-Gel™ of sizes 2 and 2.5 were used in the study. PLMA was inserted in 26 children in first attempt and 4 children in the second attempt, while I-Gel™ was inserted in 27 patients in first and 3 in the second attempt. | Table 1: Characteristics of 70 patients undergoing surgery with either ProSeal laryngeal mask airway or I-Gel™
Click here to view |
Neck flexion resulted in significant increase in the OPLP with both PLMA (P< 0.0001) and I-Gel™ (P< 0.0001). Significant reduction in OPLP was seen in neck extension with both the devices (P< 0.0001) [Table 2]. However, the OPLP was comparable with both the study devices in the three head-and-neck positions [Table 2]. The percentage (%) change in OPLP from neutral to flexion with PLMA and I-Gel™ was 25% and 17.4%, respectively (P = 0.042), while the % change in OPLP from neutral to an extension was 14.9% with PLMA and 11.8% with I-Gel™ (P = 0.39) [Table 3]. | Table 2: Comparison of oropharangeal leak pressures, peak inspiratory pressures and expired tidal volume in patients with ProSeal laryngeal mask airway and I-Gel™ in different head-and-neck positions
Click here to view |
 | Table 3: Percentage change in oropharyngeal leaking pressures, peak inspiratory pressure and expired tidal volume with change in positions from neutral with ProSeal laryngeal mask airway and I-Gel™
Click here to view |
Significantly higher PIP was recorded with I-Gel™ than PLMA during maximum flexion (17.7[4] vs. 14.6 [2.4] cm H2O, P = 0.001), [Table 2]. The percentage increase in PIP from neutral to flexion was significantly more with I-Gel™ compared to PLMA (51.2% vs. 25.9%, P < 0.001). The % change in mean PIP from neutral to an extension with PLMA and I-Gel™ was 6.1% and 7.8%, respectively [Table 3].
PLMA had significantly higher expired tidal volume inflexion compared to I-Gel™ (P = 0.0017) [Table 2]. Significantly greater percentage decrease in expired tidal volume was seen with I-Gel™ compared to PLMA (38.6% vs. 23.6%) during flexion (P = 0.001).
There was worsening of ventilation score median (range) with both PLMA 1 (0–3) and I-Gel™ 1 (0–3) inflexion of the head-and-neck. However, the ventilation scores were comparable between the two devices in all the three positions.
No adverse event with respect to displacement, desaturation, bronchospasm or laryngospasm was noted with either of the study devices.
Discussion | |  |
Although no significant difference was seen in mean OPLP in the three different head-and-neck positions, ProSeal LMA demonstrated higher percentage increase in OPLP from neutral to flexion compared to I-Gel™. A significant increase in PIP and decrease in expired tidal volume were noted during neck flexion with I-Gel™ compared to PLMA. The ventilation scores and the fibreoptic gradings were similar with both I-Gel™ and PLMA in all the three different neck positions.
The mean values of OPLP of PLMA and I-Gel™ were comparable in all the different head-and-neck positions. So far, there have been contradictory reports with regard to the comparison of OPLP between these two devices.[8],[9]
OPLP increased with flexion of the neck and decreased in neck extension with both the devices. Similar results have been replicated in studies using different SAD's in both adults and children.[10],[11],[12],[13] Jain et al. showed a significant increase in OPLP during flexion and decreased during extension with I-Gel™ in 30 anaesthetised paralysed children.[11] Similarly, an increase in OPLP with classic LMA (CLMA) during flexion and decrease during extension was observed in 39 non-paralysed anaesthetised children.[10] Flexion results in decrease in the longitudinal tension on anterior pharyngeal muscles which settle over the mask of the SAD's providing better seal and reverse is seen during extension.[14] So far, the effect of the different head-and-neck positions on OPLP of PLMA in children has not been evaluated in children. Therefore, no direct comparison can be made.
In contrast to the previous studies, we compared the mean difference and the precent change in the OPLP with each position change. In the current study, PLMA group resulted in 25% increase in the mean OPLP compared to 17.4% with I-Gel™ from neutral to flexed neck position (P = 0.042). The greater increase in mean OPLP with PLMA in comparison to I-Gel™ can be attributed to the larger and more compliable mask of PLMA compared to I-Gel™.
In another similar study comparing laryngeal tube (LT) to CLMA in children, Biedler et al. demonstrated higher OPLP with LT compared to CLMA in all the different head-and-neck positions.[15] In spite of higher OPLP with LT compared to CLMA in all the different head-and-neck positions, approximately 27% increase in OPLP with CLMA compared to 14% increase with LT with a change in position from neutral to neck flexion was neither noted or commented on. These findings further suggest the need for interpreting the results with a mean difference or percentage change with every position change, especially when there is a difference in mean values of OPLP and other ventilation parameters in the neutral position itself. This additional knowledge can alter the clinical interpretation of the results.
Neck flexion results in narrowing of the laryngeal inlet, increase in deflection of the posterior epiglottis and worsening of alignment of the pharyngeal and the laryngeal axis. Conglomeration of all these factors leads to an increase in the PIP during neck flexion.[11],[12],[13],[14],[15],[16] Along with it, compression of the cuff against a narrowed laryngeal inlet results in loss of the delivered tidal volume which was seen with both the devices during flexion.
PLMA resulted in significantly lower PIP and higher expired tidal volume in flexion compared to I-Gel™. I-Gel™ resulted in 51.2% increase in PIP and 38.6% decrease in expired tidal volume compared to 25.9% increase in PIP and 23.6% decrease in expired tidal volume with PLMA. This better preservation of ventilation parameters with PLMA can be attributed to the larger area of cross-section of the ventilation hole of PLMA (4 × 2 for size 2.5 and 3.2 × 1.5 for size 2) compared to I-Gel™ (2 × 1.8 for size 2.5 and 1.8 × 1.5 for size 2) [Figure 2] which provides PLMA the added advantage of greater buffer area for ventilation during flexion, when the laryngeal space is less, and cuff of devices are compressed. | Figure 2: Comparison of ventilation apertures of ProSeal laryngeal mask airway (4 × 2 for size 2.5 and 3.2 × 1.5 for size 2) and I-Gel™ (1.8 × 1.5 for size 2 and 2 × 1.8 for size 2.5)
Click here to view |
The greater increase in OPLP and relatively lesser change in PIP and expired tidal volume during flexion with PLMA can prove advantageous over I-Gel™ while using in surgeries requiring flexed position. As we did not perform any power calculation for the secondary outcomes, these results can be considered exploratory, and they can provide the basis of future research.
Unlike adults, the epiglottis of children is large, floppy, and more horizontally placed. In spite of the optimal placement of the supraglottic airway, a part of the epiglottis is generally visible on the fibreoptic assessment of the glottic aperture through the SAD. Therefore, in our study, we used the fiberoptic grading proposed by Okuda et al., which is better suited for the assessment of the paediatric airway.[10] Fibreoptic grades of glottis view through PLMA and I-Gel™ were comparable in all the three different positions. Most of the previous studies have shown little or no correlation between fibreoptic view and function of SADs. The changes in fibreoptic grade are considered to be a poor marker of the adequacy of ventilation.[16],[17] A study reported adequate ventilation with even poor fiberoptic score of 1–3.[18] Others have suggested radiological examination (magnetic resonance imaging) to confirm the airway location and to determine the exact site of obstruction, if present. However, such studies have not been done so far.[19]
The results of our study should be interpreted in the light of some limitations. This study was carried on anaesthetised paralysed children, and therefore, these results cannot be extrapolated to spontaneously breathing children. Flexion results in an increase in the PIP and use of muscle relaxants tend to improve the lung compliance and provide a safety margin for ventilation in a flexed position. Second, radiological confirmation for airway changes could have added to the best of our knowledge but was not feasible in our setting. In the present trial, although we had selected only size 2 and 2.5, excluding the infants and the bigger children, the difference in performance at the two ends of the range of weight of children in the study groups and variation in PIP due to the difference in the respiratory rate among different age groups cannot be ruled out. We evaluated the devices in children above 10 kg, owing to increase in the complication rate with device displacement in the younger age group. Efficacy of these devices is smaller age groups needs to be determined through future trials.
Conclusion | |  |
Oropropharyngeal leak pressure was similar with both thePLMA and I-Gel™ in the neutral position, maximum neck flexion and maximum neck extension. Although ventilation was possible with both the devices in all three positions, I-Gel™ resulted in higher PIP and lower expired tidal volume than PLMA in maximum flexion which warrants caution and future evaluation.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
References | |  |
1. | Keller C, Brimacombe J. The influence of head and neck position on oropharyngeal leak pressure and cuff position with the flexible and the standard laryngeal mask airway. Anesth Analg 1999;88:913-6. |
2. | Buckham M, Brooker M, Brimacombe J, Keller C. A comparison of the reinforced and standard laryngeal mask airway: Ease of insertion and the influence of head and neck position on oropharyngeal leak pressure and intracuff pressure. Anaesth Intensive Care 1999;27:628-31. |
3. | Park SH, Han SH, Do SH, Kim JW, Kim JH. The influence of head and neck position on the oropharyngeal leak pressure and cuff position of three supraglottic airway devices. Anesth Analg 2009;108:112-7.  [ PUBMED] |
4. | Cook T, Howes B. Supraglottic airway devices: Recent advances. Contin Educ Anaesth Crit Care Pain 2011;11:56-61. |
5. | Intersurgical. I-gel™ User Guide. Wokingham, Berkshire: Intersurgical; 2005. |
6. | Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth 1999;82:286-7. |
7. | Berry AM, Brimacombe JR, McManus KF, Goldblatt M. An evaluation of the factors influencing selection of the optimal size of laryngeal mask airway in normal adults. Anaesthesia 1998;53:565-70. |
8. | Nirupa R, Gombar S, Ahuja V, Sharma P. A randomised trial to compare I-gel and ProSeal™ laryngeal mask airway for airway management in paediatric patients. Indian J Anaesth 2016;60:726-31.  [ PUBMED] [Full text] |
9. | Saran S, Mishra SK, Badhe AS, Vasudevan A, Elakkumanan LB, Mishra G, et al. Comparison of I-gel supraglottic airway and LMA-ProSeal™ in pediatric patients under controlled ventilation. J Anaesthesiol Clin Pharmacol 2014;30:195-8.  [ PUBMED] [Full text] |
10. | Okuda K, Inagawa G, Miwa T, Hiroki K. Influence of head and neck position on cuff position and oropharyngeal sealing pressure with the laryngeal mask airway in children. Br J Anaesth 2001;86:122-4. |
11. | Jain D, Ghai B, Bala I, Gandhi K, Banerjee G. Evaluation of I-gel™ airway in different head and neck positions in anesthetized paralyzed children. Paediatr Anaesth 2015;25:1248-53. |
12. | Brimacombe J, Keller C. Stability of the LMA-ProSeal and standard laryngeal mask airway in different head and neck positions: A randomized crossover study. Eur J Anaesthesiol 2003;20:65-9. |
13. | Sanuki T, Uda R, Sugioka S, Daigo E, Son H, Akatsuka M, et al. The influence of head and neck position on ventilation with the I-gel airway in paralysed, anaesthetised patients. Eur J Anaesthesiol 2011;28:597-9. |
14. | Isserles SA, Rozenberg B. LMA – reduction of gas leak. Can J Anaesth 1995;42:449. |
15. | Biedler A, Wrobel M, Schneider S, Soltész S, Ziegeler S, Grundmann U, et al. Randomized evaluation of the size 2 laryngeal tube and classical laryngeal mask airway in different head and neck positions in children under positive pressure ventilation. J Anesth 2013;27:657-62. |
16. | Inagawa G, Okuda K, Miwa T, Hiroki K. Higher airway seal does not imply adequate positioning of laryngeal mask airways in paediatric patients. Paediatr Anaesth 2002;12:322-6. |
17. | van Zundert A, Brimacombe J, Kamphuis R, Haanschoten M. The anatomical position of three extraglottic airway devices in patients with clear airways. Anaesthesia 2006;61:891-5. |
18. | Xue FS, Mao P, Liu HP, Yang QY, Li CW, He N, et al. The effects of head flexion on airway seal, quality of ventilation and nasogastric tube placement using ProSeal ™ laryngeal mask airway. Anaesthesia 2008;63:979-85. |
19. | Nandi PR, Charlesworth CH, Taylor SJ, Nunn JF, Doré CJ. Effect of general anaesthesia on the pharynx. Br J Anaesth 1991;66:157-62. |
[Figure 1], [Figure 2]
[Table 1], [Table 2], [Table 3]
|