LETTER TO EDITOR
Year : 2015  |  Volume : 59  |  Issue : 12  |  Page : 827-829  

Gum elastic bougie as a tube exchanger: Modified technique


1 Department of Anaesthesia, Dharwad Institute of Mental Health and Neurosciences, Dharwad, Karnataka, India
2 Department of Anaesthesia, Karnataka Institute of Medical Sciences, Hubli, Karnataka, India

Date of Web Publication11-Dec-2015

Correspondence Address:
Goneppanavar Umesh
DIMHANS, Dharwad, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.171604

Rights and Permissions

How to cite this article:
Umesh G, Sushma K S, Sindhupriya M, Swarnamba U N. Gum elastic bougie as a tube exchanger: Modified technique. Indian J Anaesth 2015;59:827-9

How to cite this URL:
Umesh G, Sushma K S, Sindhupriya M, Swarnamba U N. Gum elastic bougie as a tube exchanger: Modified technique. Indian J Anaesth [serial online] 2015 [cited 2020 Nov 26];59:827-9. Available from: https://www.ijaweb.org/text.asp?2015/59/12/827/171604

Sir,

The standard gum elastic bougie (GEB) which is used mainly for aiding in difficult laryngoscopy and intubation can be used as a tube exchanger also. The essential steps of tube exchange using GEB require it to be inserted into the lumen of the endotracheal tube (ETT) until the markings of the GEB correlate with the markings on the ETT. This indicates that the tip of the GEB is at the tip of the ETT. [1] Subsequently, cuff of the ETT is deflated and the ETT withdrawn gently over the GEB, whereas the operator end of the GEB is held securely. However, by the time, the tip of the ETT comes to lie in the oral cavity; the whole length of the GEB gets lost within the ETT lumen and the oral cavity resulting in the anaesthesiologist losing control over the GEB completely. This is because the GEB measures 60 cm in length and an ETT of 8.0 mm inner diameter (ID) measures 36 cm [from its distal end to the edge of the blue connector, [Figure 1]a, thus resulting in only 24 cm of GEB length remaining outside the ETT. Furthermore, the ETT is usually fixed around 22-23 cm at the angle of the mouth. Hence, overall 59-60 cm length of the GEB is required to be inserted into the ETT (36 cm ETT length and 22-23 cm length inside the oral cavity) before the shaft of the GEB could be recovered outside the oral cavity, when the ETT tip comes out. Therefore, invariably the anaesthesiologist ends up losing control over the GEB for a brief while. This total loss of control over the GEB can result in the GEB either migrating deeper into the trachea, thus stimulating the carina in an already/almost awake patient, or there is a risk that the GEB comes out of the larynx along with the ETT and gets accidentally placed in the oral cavity or oesophagus. The traditional airway exchange catheters measure more than 80 cm (83 cm, 11 and 14 G), and, therefore, do not have this problem. [2] As long as the length of the GEB is more than twice that of the ETT, this problem should not happen. Bougies from other manufacturers that measure approximately 70 cm in length are available, and these problems are unlikely to be encountered with these. [3]
Figure 1: (a) Length of the 8.0 mm internal diameter endotracheal tube is seen to be approximately 36 cm, and when gum elastic bougie is inserted appropriately, 24-25 cm of it projects outside the endotracheal tube. The figure also shows a shortened endotracheal tube of 8.0 mm internal diameter, which is cut at 24 cm mark. This allows for nearly 36 cm of the gum elastic bougie projecting out of the endotracheal tube. (b) A 4.5 mm internal diameter uncuffed endotracheal tube snugly fits the distal end of the gum elastic bougie thus lengthening the gum elastic bougie to >80 cm

Click here to view


Having experienced these problems with the traditional GEB when used as tube exchanger on a couple of occasions, we have devised two unique solutions to overcome these problems.

First, prior to planned extubation of the ETT, the ETT is cut at its 24-26 cm mark (1 or 2 cm above the oral cavity), and the blue connector is fixed [Figure 1]a. Since the ETT length now is reduced by almost 10 cm relative to ETT, it ensures that the length of GEB (60 cm) is now more than double that of the ETT (26 × 2 = 52 cm). Hence, nearly 34-36 cm of it lies outside the ETT when correctly positioned and therefore, it allows the anaesthesiologist to be always in control of the GEB during the procedure of extubation. However, it requires that another ETT be cut to similar length and kept beside the patient for future tube exchange if need be. The second solution involves attaching an uncuffed 4.5 mm ID ETT to the proximal end of the bougie. This snugly fits the GEB [Figure 1]b and requires considerable force to separate the two. Thus, this setup helps in lengthening the GEB to >80 cm.

We are confident these solutions to GEB can help minimise the complications discussed earlier. While a GEB may not be suitable for tube exchange through the nasal route due to the limitation of length, these 2 modifications can enable the traditional GEB to be used through both oral and nasal routes. Though an airway exchange catheter is ideal for extubation of the difficult airway and tube exchange, the modified GEB with the second solution that we have provided appears to have an edge as it is useful in aiding difficult intubation also.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Cavallone LF, Vannucci A. Review article: Extubation of the difficult airway and extubation failure. Anesth Analg 2013;116:368-83.  Back to cited text no. 1
    
2.
Difficult Airway Society Extubation Guidelines Group, Popat M, Mitchell V, Dravid R, Patel A, Swampillai C, et al. Difficult Airway Society Guidelines for the management of tracheal extubation. Anaesthesia 2012;67:318-40.  Back to cited text no. 2
[PUBMED]    
3.
Available from: http://www.smiths-medical.com/plugins/news/2006/mar/new-portex-single-use.html. [Last accessed on 2015 Jun 16].  Back to cited text no. 3
    


    Figures

  [Figure 1]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed1964    
    Printed8    
    Emailed0    
    PDF Downloaded352    
    Comments [Add]    

Recommend this journal