Year : 2011  |  Volume : 55  |  Issue : 1  |  Page : 83-84  

Yet another modification of Guedel's airway


1 Department of Anaesthesia & Intensive Care, Government Medical College and Hospital, Sector-32, Chandigarh, India
2 Department of Anaesthesia & Intensive Care, Postgraduate Institute of Medical Education and Research, Chandigarh, India

Date of Web Publication9-Feb-2011

Correspondence Address:
Vanita Ahuja
813, Sector-7, Panchkula - 134 114
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0019-5049.76579

Rights and Permissions

How to cite this article:
Ahuja V, Arya VK, Kumar B. Yet another modification of Guedel's airway. Indian J Anaesth 2011;55:83-4

How to cite this URL:
Ahuja V, Arya VK, Kumar B. Yet another modification of Guedel's airway. Indian J Anaesth [serial online] 2011 [cited 2020 Nov 26];55:83-4. Available from: https://www.ijaweb.org/text.asp?2011/55/1/83/76579

Sir,

Flexible fibreoptic laryngoscopic intubation (FFLI) is a very reliable approach to difficult airway management. [1] Oral FFLI is considered more difficult than nasal FFLI as the tip enters the larynx at an acute angle to the glottis. [1] In anaesthetised patients, there is loss of tone in the submandibular muscles, tongue and indirectly the epiglottis. Flexible fibreoptic laryngoscope (FFL) is expensive, delicate, and the cables are not strong enough to lift or dislodge the tissues. [1] A number of fibreoptic-compatible oral airways (FCOAs) have been designed to protect the instrument from the patient's teeth, guide it into the midline and keep the tongue from falling backwards.[2]

We have devised a modification of Guedel's airway by removing the palatal surface of the airway [Figure 1]. The tracheal tube is threaded onto the fibrescope before laryngoscopy. The modified airway keeps the FFL tip around the back of the tongue, in the midline position and close to the larynx till it enters the trachea. The airway can be removed with a lateral twist movement without disturbing the endotracheal tube-endoscope assembly. Following this, the tracheal tube is advanced and the FFL is withdrawn.
Figure 1: Guedel's airway with slit on the palatal surface of the airway

Click here to view


On comparing it with anterior (lingual) channel like the Patil-Syracuse and the Williams airway intubator, the latter have the advantage of better localisation of the glottis opening but need to be removed from the oropharynx before a tracheal tube can be advanced over the fibre optic cable into the glottis, creating an unnecessary extra step. [1] McGinley and McAdoo [3] devised a modification of Guedel's airway with deficiency in anterior support. However, at times, the tip of the FFL gets lodged in the pyriform fossa due to the deficiency in anterior support. Oral airways with a posterior channel, such as the Ovassapian and Berman, facilitate easy fibreoptic oro-tracheal intubation. Each can be rapidly removed from around the tracheal tube except for Burman's airway as it has both posterior and lateral channels. [1] We have done a pilot study comparing both lingual and palatal modifications of Guedel's airway and found that the palatal modification of Guedel's airway makes FFLI easy, keeps the FFL in midline, no addition manoeuver like jaw thrust is required and causes no disturbance of endotracheal tube-laryngoscope assembly. A possible limitation of the modification would be lack of standardisation. Geudel's airway is available in many sizes as compared to other FCOAs, making it versatile, cost-effective and highly acceptable.

 
   References Top

1.Dorsch JA, Dorsch SE. Face masks and airways. In: Dorsch JA, Dorsch SE, editors. Understanding Anesthesia Equipment. 5th ed. Baltimore, USA: Williams and Wilkins; 2008. p. 443-59.  Back to cited text no. 1
    
2.Gal TJ. Airway management. In: Miller RD, editor. Miller's anesthesia. 6th ed. Philadelphia, PA: Elsevier Churchill Livingstone; 2005. p. 1617-52.   Back to cited text no. 2
    
3.McGinley J, McAdoo J. Airway adjunct to an unanticipated difficult airway. Anesth Analg 1999;88:467-8.  Back to cited text no. 3
    


    Figures

  [Figure 1]


This article has been cited by
1 One airway many modifications
Kumar, P., Johar, S., Bajaj, M., Hooda, S.
Indian Journal of Anaesthesia. 2011; 55(5): 552
[Pubmed]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    References
    Article Figures

 Article Access Statistics
    Viewed2040    
    Printed51    
    Emailed0    
    PDF Downloaded349    
    Comments [Add]    
    Cited by others 1    

Recommend this journal